Calibration & Preventative Maintenance

0
163

Calibration & Preventative Maintenance Procedures :

Calibration: A comparison of two instruments or measuring devices one of which is a standard of known accuracy (traceable to national standards) to detect, correlate, report or eliminate by adjustment, any discrepancy in accuracy of the instrument measuring device being compared to the standard.

Calibration Programs Required by Regulatory Authorities

• “Automatic, mechanical, or electronic equipment or other types of equipment, including computers, or related systems that will perform a function satisfactorily, may be used in the industries like power plant, oil & gas, cement, manufacture, processing, packing etc. If such equipment is so used, it shall be routinely calibrated, inspected, or checked according to a written program designed to assure proper performance.

• Maintenance at appropriate intervals to prevent malfunction & shall be “preventative” not “reactive” maintenance.

• Calibration requirements for Laboratory Instruments

–Specific Directions
–Schedules
–Limits of accuracy & precision
–Remedial Actions
–Systems to prevent usage of instruments failing calibration

• “Control, weighing, measuring, monitoring and test equipment that is critical for assuring the quality of intermediates or APIs should be calibrated according to written procedures and an established schedule.

Each Manufacturing / Process Area:

• Written calibration procedures that use traceable calibration standards or calibration equipment.

• Preventive maintenance procedures and / or referenced manuals

• Qualified individuals (having the appropriate education, training, background and experience) responsible for calibrating & maintaining instrumentation

• Second person check of all calibration and maintenance

• Qualified individuals responsible for monitoring the calibration and maintenance program.

• Ensure the calibration program and procedures are reviewed and approved by Quality

Instrument / Equipment Master List

• System for identification of all Master equipment related to instrumentation in a manufacturing/process area or laboratory

• Include instrumentation details (serial number, model number & location)

• If automation components are tracked separately through the configuration management then it is not necessary to include (must verify)

• Procedures must exist that identify the calibration and maintenance requirements for each instrumentation / equipment on the master list

Retired Equipment:

–Records pertaining to retired / obsolete equipment must be kept according to company’s records retention procedures

–Records should include date the unit was retired, person responsible and the reason for retirement / discard

Instrument Identification & Calibration Status

• Each instrument given a unique identifier

• Instrumentation details associated with this number must be documented and available (e.g. serial number, model number, location, etc.)

• Each instrument should be labeled with the unique identifier

• Calibration status of each instrument , the date of calibration, the next calibration date and the identification of person performing calibration should be readily available

•Appropriate systems to document calibration status include calibration logs, MAXIMO, and calibration stickers

• System must be in place to prevent use of an instrument that is not qualified, unusable due to damage or malfunction, or has exceeded its established calibration interval

• System must be in place that identifies instruments that do not require calibration to be performed beyond the original or factory calibration to distinguish from those instruments that do require scheduled calibrations

⇒Documentation required for excluding equipment

Traceability of Standards / Calibration Equipment

• Calibration reference standards / calibration equipment shall be traceable to national standards and be accompanied by certificates of traceability / analysis

• If recognized standards are not available, an independent reproducible standard may be used

• The calibration tolerance of a given standard should be as tight or tighter than the tolerance of the instrument to be calibrated

•A procedure must be in place to ensure tracking and monitoring of standard’s expiration date and re-calibration / re-certification requirements

• Re-calibration records must be retained

Instrument Calibration Tolerances

•Instrument calibration tolerance limits should be established so problems are identified and corrected in a timely manner

•When assigning tolerances, considerations given to:

  • Capability of the instrument being calibrated (what the manufacturer/OEM claims the instrument can achieve).
  • Parameters at which the instrument operates (ex: if testing accuracy of + 0.5% is required, the instrument calibration tolerances should be <0.5%)
  • Work environment – environmental conditions can affect the performance of the instrumentation

•Practice of using “Alert” & “Action” levels

•“Alert” Tolerance (“Adjustment Limit”)

  • Related to instrument performance
  • Level at which the instrument is adjusted back into range
  • Not required – an industry “best practice”

•“Action” Tolerance (“Calibration Limit” or “Out-of-Tolerance”)

  • Tied to process performance
  • Level at which the potential for product impact exists
  • Possible atypical investigation or reporting is required

•Deviations beyond “Alert” level but not at “Action” level would not require investigation – May require adjustment, changes to PMs

•Setting of “Alert” and “Action” levels should be described in SOPs, be defendable and have Quality review and approval

Calibration and Maintenance Frequencies

•May be determined for individual instruments or groups of instruments (similarity of construction, reliability, and stability)

•Considerations when determining calibration frequency:

  • Accuracy of the measurement / instrument range
  • Consequences of incorrect value caused by out of calibration
  • Extent & criticality of use of the instrument & tendency to wear and drift
  • The manufacturer’s/OEM’s recommendations
  • Environmental & physical conditions (temperature, humidity, vibration)
  • Previous calibration records, history of calibration problems & repair history
  • Frequency of calibration checks prior to use or in-between intervals
  • Redundant / back-up systems (provides secondary source of information available from other calibrated primary instruments)i
  • Results of Qualification studies
  • Process requirements
  • Availability of built-in / automatic calibration checks

• Changes to frequency must be approved per change control SOPs

• References to specific instrument procedures listed in compendia (USP, BP or EP) for calibration tests for specific laboratory instruments

• Calibration time “windows” should be established around calibration due dates

• Policy for Calibration and Maintenance Intervals & Schedules Include:

  • Extending Intervals
  • Reducing Intervals
  • Maximum Intervals

Maintenance Requirements

•Manufacturer’s/OEM’s recommendations

• Parts that wear: gaskets, seals & bearings

• Parts requiring periodic replacement: filters, belts & fluids

• Parts requiring periodic inspection and cleaning

• Parts requiring periodic adjustments, tightening and lubrication

Calibration and Maintenance Procedures

• Shall include specific directions and limits for accuracy & precision

 

• Shall include guidance for remedial action when accuracy & precision limits are not met

• Normally provided in the manufacturer’s/OEM’s manuals

• Some compendial requirements exist for some specific laboratory instrumentation

• Performance checks (e.g. system suitability, daily balance performance checks) are NOT suitable substitutes for regularly scheduled calibrations

Each calibration & maintenance procedure should include the following:

  • Identification of department responsible to perform the calibration or maintenance
  • Step-by-step calibration instructions, reference to appropriate calibration procedures or instrument manuals
  • Methods for preventive maintenance or reference to appropriate instrumentation manuals
  • Calibration equipment used in the calibration (e.g. spectroscopy filters, voltmeters, digital thermometers, etc)
  • Calibration parameter and tolerance ( ± )

Each calibration & maintenance procedure should include the following:

Required environmental controls or conditions, where appropriate

  • Provisions for adjustments, if needed
  • Requirement for recording actual measurements before (“as found”) and after adjustment or preventive maintenance (“as left”)
  • Actions to be taken if instrumentation cannot be calibrated (e.g. contact appropriate service people, label and remove from service)
  • A step to record all calibration & maintenance activities.

Note : “as found” recordings are not required where routine performance checks are available to provide evidence indicating instrument is operating properly and is suitable for use.

Use of Contractors / Vendor Service Personnel

– Must have a procedure for approving contractor activities

– Contractors must have appropriate training / qualification

– Responsible to review & approve the contractors calibration and maintenance procedures.

– Responsible to perform actions or steps not part of the contractors procedures.

Calibration & Maintenance Records

– All calibration records must be retained per document retention procedures

– Should include “as found” measurements, results of adjustments (“as left”) and appropriate review & approval of all results

– Tolerance or limit for each calibration point

– Identification of standard or test instrument used

– Identification of persons performing the work and checking the results with dates

– Review must ensure the approved activities have been completed and all results have passed the established acceptance criteria

– Periodic review of historic calibration & maintenance data to evaluate appropriateness of established frequencies